SKU:
ATMOS-GAS-CELL
MPN:
GS24654
Stock Status:
In Stock
Introduction
Specac’s standard long pathlength gas cell range is now the new Atmos™ series of metal-bodied fixed-pathlength gas cells. They have been optimized using the latest optical modelling tools to give maximum signal-to-noise.
The Atmos™ range of cells are heatable to 200 °C with the addition of a heating jacket and temperature controller.
Features
- Optimized optical path for unmatched signal-to-noise
- Low cell volume for fast gas exchange
- Ni-coated Al cell bodies for improved thermal transfer from heating jackets. Also holds higher gas pressure than glass.
- Inert stainless-steel mirror substrates and avoidance of glues to prevent outgassing
- 125 psi (8.6 bar) maximum cell pressure
- Retrofittable heating jacket option for up to 200 °C
Applications
Gas detection, measurement, FTIR
Order Information
Model | A2.5 | A5 | A10 | A20 |
---|---|---|---|---|
Pathlength (m) | 2.5 | 5.0 | 10 | 20 |
Pass pathlength (mm) | 104 | 139 | 250 | 455 |
Number of passes | 24 | 36 | 40 | 44 |
Volume (L) | 0.27 | 0.63 | 2.12 | 3.68 |
Pathlength Selection Guide
Gas name | formula | v (cm-1) | ppm.m | absorbance |
Carbon dioxide | CO2 | 2360 | 100 | 0.40 |
Carbon monoxide | CO | 2170 | 100 | 0.04 |
Methane | CH4 | 3020 | 100 | 0.10 |
C2 to C6 n-alkanes | 2960 | 100 | 0.10 | |
Nitrogen dioxide | NO2 | 1630 | 100 | 0.15 |
Nitric oxide | NO | 1900 | 100 | 0.015 |
Sulfur dioxide | SO2 | 1370 | 100 | 0.09 |
Hydrogen sulfide | H2S | 1300 | 1000 | 0.002 |
Ammonia | NH3 | 960 | 100 | 0.12 |
Hydrogen chloride | HCl | 2940 | 100 | 0.04 |
Water | H2O | 1650 | 1000 | 0.20 |
Vinyl chloride | CH2CHCl | 950, 900 | 100 | 0.06 |
Acetaldehyde | CH3CHO | 2750 | 100 | 0.015 |
Benzene | C6H6 | 670 | 10 | 0.09 |
Toluene | C6H5CH3 | 730, 690 | 100 | 0.10 |
Methanol | CH3OH | 1040 | 100 | 0.10 |
Ethanol | CH3CH2OH | 1050 | 100 | 0.05 |
Carbonyl sulfide | COS | 2070 | 100 | 0.40 |
Nitrous oxide | N2O | 2235 | 100 | 0.15 |
Sulfur hexafluoride | SF6 | 950 | 10 | 0.40 |