Calvet Pro (SENSYS) Simultaneous TG-DSC or TG-DTA (STA)
Calvet Pro (SENSYS) Simultaneous TG-DSC or TG-DTA (STA)
Information
Differential Scanning Calorimeter- Incomparable precision: the unique 3D DSC sensor totally surrounds your sample and reference crucibles so that the entire energy of any transformation is monitored thereby giving you an unequalled accuracy of measurement.
Features
‧CRUCIBLES Crucibles of 150 μl capacity are available in aluminium, alumina, platinum or silica to suit the chemical nature of the sample. A silica tube can be added to protect the transducer and the analysis chambers during test involving sweeping with corrosive gas.
‧Evolved Gas Analysis The SENSYS evo design features both a small furnace volume and also an easily accessible exhaust to allow for direct connection to further analysis devices such as MS, FTIR, GC etc
‧Measurements under controlled humidity Humidity is known to have a significant influence on the stability of many materials.
‧SENSYS evo TG-DSC as well as other analysis instrument(RX, GC, etc.) for studies of sorption, hydration, stability, etc.
Applications
pharmaceuticals (phase diagram, polymorphism, purity, thermal stability), safety of chemical process (transition, decomposition under high pressure), energy (catalysis, hydrogen storage, hydrogen adsorption),polymers (glass transition, measurement under pressure), etc. View the application notes in your field, available for download,
Specifications | |
---|---|
[Temperature range]: | -120~830℃ |
[Scanning rate]: | 0.01~30 K/min |
[Maximum balance capacity]: | 3g |
[Weight range]: | +/-200 mg |
[TG resolution]: | 0.03μg |
[Baseline dynamic drift]: | < 10 μg |
[DSC Resolution]: | 0.35μW/0.035μW |
[Autosampler]: | 48 samples even under pressure |
[Gases]: | 3 carrier gases (MFC from 4 to 200 ml/min) + 1 auxiliary or reactive gas (MFC from 0.3 to 16 ml/min) |
[Cooling system]: | Liquid Nitrogen Accessory |